Télécharger Imprimer la page

GERATHERM smart GT-1775 Mode D'emploi page 72

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 76
Appendix
Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
and/or user of the device should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
EN
NOTE 1:
NOTE 2:
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the device.
b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IEC 60601 –
Compatibility
test level
level
150 kHz to
150 kHz to
80 MHz:
80 MHz:
3 Vrms
3 Vrms
6 Vrms (in ISM
6 Vrms (in ISM
and amateur
and amateur
radio bands)
radio bands)
80 % Am
80 % Am
at 1 kHz
at 1 kHz
10 V/m,
10 V/m,
80 % Am
80 % Am
at 1 kHz
at 1 kHz
At 80 MHz and 800 MHz the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
Notes concerning the electromagnetic
environment
Portable and RF communications equipment
should be used no closer to any part of the
device, including cables, than the recom-
mended separation distance calculated from
the equation appropriate for the frequency of
the transmitter.
Recommended separation distance:
d = 0,5
d = 1,2
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in metres (m).
80 MHz
Field strengths from fixed RF
to
transmitters, as determined
800 MHz:
by an electromagnetic site
d = 1,2
survey, should be less than
the compliance level in each
800 MHz
frequency range.
to
2,7 GHz:
Interference may occur in the
d = 2,3
vicinity of equipment marked
with the following symbol:
72

Publicité

loading