Appendix
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer
and/or user of the device should assure that it is used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker
emissions
IEC 61000-3-3
Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer
and/or user of the device should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / Burst
EN
IEC 61000-4-4
Surge voltage
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50 Hz / 60Hz)
Magnetic field
IEC 61000-4-8
Compatibility
Group 1
Class B
Class A
Compatible
IEC 60601 -
Compatibility
Test level
level
± 8 kV contact
± 8 kV contact
± 15 kV air
± 15 kV air
Power supply
lines:
Power supply
± 2 kV
lines:
Input/output lines:
± 2 kV
± 1 kV
Line(s) to line(s):
± 1 kV
Line(s) to line(s):
Line(s) to earth/
± 1 kV
ground:
100 kHz repetition
± 2 kV
frequency
100 kHz repetition
frequency
0 % 0.5 cycle at
0 % 0.5 cycle at
0°, 45°, 90°,
0°, 45°, 90°,
135°, 180°, 225°,
135°, 180°, 225°,
270° and 315°
270° and 315°
0% 1 cycle and
0 % 1 cycle and
70 % 25/30
70 % 25/30 cycles
cycles
Single-phase: at 0
Single-phase: at 0
0 % 300 cycles
0 % 300 cycles
30 A/m
30 A/m
50 Hz / 60 Hz
50 Hz / 60 Hz
70
Notes concerning the electromagnetic
environment
RF energy is used only to maintain the device's
operation. Therefore, its RF emissions are so
low that it is not likely to cause any interference
in nearby electronic equipment.
The device is suitable for use in all establish-
ments, other than domestic and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Notes concerning the electromag-
netic environment
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.