Supplement SW 1.04.x4
Alarm (ID)
Phase/Priority/Alarm mute time [sec.]
Final conductivity limit (1029)
Pre / low / 0
The / low / 180
Reinf / low / 0
Final conductivity limit (SUP)
(1951)
Pre / low / 0
The / low / 120
Reinf / low / 120
UF volume exceeded (SUP)
(1966)
The / low / 120
Reinf / low / 120
bioLogic plausibility check failed
(SUP) (1982)
The / low / 120
EN
Supplement to Dialog iQ Instructions for Use: SW 1.04.x4
Read the following information and instruct all users in your organization.
Insert this supplement into all Dialog iQ Instructions for Use, Revision
1.04.01, when software 1.04.x4 is installed at your machine.
This supplement only contains information that are changed. For the
complete descriptions, refer to the existing Dialog iQ Instructions for Use.
With software version 1.04.x4, two limit values have been changed. An
additional bioLogic alarm (1982) has been implemented and the information
to alarm 1966 has been changed. The information to expiration of filters
(2078, 2079) have been changed. The resulting new alarm texts are:
Cause and Remedial Action
Description:
Final conductivity deviates more than 2.5% from set value.
Possible Cause:
- Wrong acid/acetate concentrate
- Air entrance before conductivity sensor
Next steps:
Check connected concentrate.
Note:
If alarm persists, contact technical service.
Description:
Supervisor Alarm: Final conductivity deviates more than 3% from set
value.
Possible Cause:
- Wrong acid/acetate concentrate
- Air entrance before conductivity sensor
Next Steps:
Check concentrate.
Note:
If alarm persists, contact technical service.
Description:
Supervisor alarm: UF volume deviation detected.
Possible Cause:
- defective UF pump
Next Steps:
- End therapy
- Check patient weight
Note:
Contact technical service.
Description:
Supervisor Alarm: Implausible calculation of bioLogic feedback controller
detected.
Next Steps:
Confirm alarm to continue bioLogic mode.
If alarm reoccurs:
- deactivate bioLogic mode to continue therapy
- check UF settings
Supplement 38910517 / Rev. 1.00.00 / 2021-12
Dialog iQ