MENUS DES SORTIES mA

MAIN MENU
Select a menu using
then press Enter
Go to: mA Outputs
Enter Key
mA Outputs
Select a menu using
then press Enter
Options:mA Output #(default), Set Hold Mode
Go to: xxxxx
mA OUTPUT#1
P-Set: A (None)
Select Signal: XXX
XXX: O2 (unit as defined in parameter set menu), Temperature
mA Output #
mA OUTPUT #1
Options: 1,2,3,4
P-Set: A (NAME)**
Type: Normal
Options:Normal, Bi-Linear AutoRange, Logarithmic
Low Value: 4mA
Options: 4 mA (default), 0 mA
AutoRange
mA OUTPUT #1
Range 1 Settings:
4mA = 0.000 (Unit)*
20mA = 0.000 (Unit)*
mA OUTPUT #1
Range 2 Settings:
4mA = 0.000 (Unit)*
20mA = 0.000 (Unit)*
mA OUTPUT #1
On failure 22mA: Yes
Current Out = 22.00mA
On failure 22mA Options:
Yes, No
*(Unit) = O2 unit defined in Parameter Set Menu
**(NAME) = Name as defined in menu parameter set.
Annexe A
Logarithmic
Bi-Linear
mA OUTPUT #1
4mA = 0.0000
(Unit)*
Mid = 0.0000
20mA = 0.0000
mA OUTPUT #1
On failure 22mA: Yes
Current Out = 22.00mA
Arborescence du menu
mA OUTPUT #1
20mA = 0.0000 (Unit)*
Num of Decades:1
SET HOLD VALUE
more
Hold state: No Hold
mA OUTPUT #1
On failure 22mA: Yes
Current Out = 22.00mA
Normal
mA OUTPUT #1
On failure 22mA:Yes
P-Set A (NAME)
4mA = 0.000 (Unit)*
Current Out = 22.00mA
20mA = 0.000 (Unit)*
Set Hold Mode
SET HOLD MODE
P-Set A (NAME)**
Hold state:xxxxx
No Hold (default)
SET HOLD FIX
P-Set A (NAME)**
P-Set A (NAME)**
Fix Value: 22.0 mA
Option: Last
mA OUTPUT #1
On failure 22mA Options:
Yes, No
Option:
Fix
Last
32