Accessories and compatible catheters length
Item
Compatible Catheters
PSO-PB/PSO-PBT
PSO-PT/PSO-PTT
PSO-VT/PSO-VTT
Operation on batteries
Item
Specification
Battery life
>6 h during valid maintenance period
Indicative
charging
time
10 h for full charge
for a new bat-
tery
Environmental specifications
Item
Specification
Operation
+10 °C (50 °F) to +40 °C (104 °F)
Temperature
Storage,
-20 °C (-4 °F) to +60 °C (140 °F)
transport
Operation
15% to 95% relative humidity without
condensation
Humidity
Storage,
5% to 95% relative humidity
transport
Operation
-500 to 3000 m
Altitude
Storage,
-500 to 4600 m
transport
Measuring and display specifications
Item
Specification
Display range
Unit
Resolution
Accuracy
ICP
Sampling
Smoothing rate 8 s
Refreshing
Systolic range
Systolic and
diastolic
Diastolic range -50 mmHg to +150 mmHg
pressures
Resolution
Unit
ICT
Display range
Resolution
Length
1 m
-40 mmHg to +150 mmHg
mmHg
+1 mmHg
+/-2% of the reading from 0 to
+100 mmHg or +/-2 mmHg
100 samples/s
1 s
-40 mmHg to +200 mmHg
1 mmHg (Sys. and Dias.)
Celsius (°C) or Fahrenheit (°F)
°F = °C x (9/5) + 32
°C = (°F - 32) x (5/9)
Equivalence in Kelvin (K):
K = °C + 273.15
K = °F x (5/9) + 255.37
+20.0 °C to +45.0 °C (+68.0 °F to
+113.0 °F)
0.1 °C (0.1 °F)
Measuring and display specifications
Item
Specification
Accuracy
Refreshing
Mode
Time constant
Externalization specifications to a patient monitor
Item
Specification
Output range
(mmHg)
Excitation voltages
Max. input voltage
Output range (Hz)
Sensitivity
ICP output
to a patient
Resolution
bedside
monitor
Accuracy
Sampling
Calibration points
Alert pressure
Refreshing
Output range
Sensitivity
Resolution
ICT output
to a patient
Accuracy
bedside
monitor
Max. input voltage
Refreshing
Externalization specifications to a PC
Item
Specification
USB
RS232
Media Format
Socket
Max. input voltage + 5 Vdc
Output range
Resolution
ICP output to a PC
Sampling
Alert pressure
Assigned output range +/- 0.2 °C
from 25.0 °C to 45.0 °C
Assigned extended output range +/-
0.4 °C from 20.0 °C to 25 °C
1 s
Direct
Less than 10 s for parenchymal
Catheters: PSO-PB, PSO-PBT, PSO-
PT or PSO-PTT
Less than 20 s for ventricular Cathe-
ters: PSO-VT or PSO-VTT
-40 mmHg to +150 mmHg
+2 to +8 Vdc or +2 to +8 Vac
+8 Vdc or +8 Vac
DC to 5000 Hz
+5 µV/V/mmHg
+0.125 mmHg
+/- 2% of the reading of 0 to
+100 mmHg or +/-2 mmHg
100 samples/s
0, +30, and +60 mmHg
+360 mmHg
10 ms
+20.0 °C to +45.0 °C (+68.0 °F to
+113.0 °F)
Standard YSI 400
0.1 °C (0.1 °F)
Assigned output range +/- 0.2 °C
from 25.0 °C to 45.0 °C
Assigned extended output range
+/- 0.4 °C from 20.0 °C to 25.0 °C
+5 Vdc
1 s
USB 2.0
115200 bauds, 1 start bit, 1
stop bit, no parity bit
Mini USB type B female
-40 mmHg to +150 mmHg
+0.1 mmHg
100 samples/s
+360 mmHg
Page 29 of 174