CAUTION
Federal (USA) law restricts this device to sale by or on
the order of a physician.
1. Figure
Figure 1. Representative Pressio
2. Introduction
These Instructions for Use detail all the information required
for the installation, use and maintenance of the Monitor. They
also provide thorough explanations of all information dis-
played on the screen.
After reading these instructions carefully, the operator will be
able to:
– Connect cables
– Connect Catheter
– Zero Catheter
3. Intended Use
®
The Pressio
2 Intracranial Pressure (ICP) monitor is de-
signed to read and display the pressure values and curves
®
measured by the Pressio
catheters implanted in patients re-
quiring continuous intracranial pressure monitoring.
Depending on the type of catheter used, the Pressio
monitor can also display the intracranial temperature.
With the appropriate Pressio
monitor also makes it possible to display the ICP curve and
the Intracranial Temperature (ICT) value on a patient monitor.
®
The Pressio
2 Intracranial Pressure (ICP) monitor is inten-
ded for hospital use, particularly in the neuro-intensive care
and neurosurgery (operating room) departments.
Patients undergoing ICP (and ICT) monitoring should be sub-
ject to strict, continuous monitoring so as to prevent any
complications from arising.
®
Monitoring System configuration
®
2 accessories, the Pressio
NOTE
Read the Instructions for Use carefully before using the
®
Pressio
2 ICP Monitor (hereinafter referred to as Moni-
tor). Do not use the Quick Start Guide before reading
the current Instructions for Use.
Refer to the Instructions for Use of the specific Pressio
Catheter being used (hereinafter referred to as Cathe-
ter): PSO-PB, PSO-PT, PSO-VT, PSO-PBT, PSO-PTT or
PSO-VTT.
[1] Power cable [2] Pressure cable [3] Temperature cable
®
[4] Pressio
Catheter [5] Catheter extension cable [6] Press-
– Monitor the patient
– Access the monitoring history
– Perform basic maintenance
NOTE
The patient is not the expected operator.
For more information, go to www.sophysa.com, or contact
Sophysa at contact@sophysa.com, or +33(0)1 69 35 35 00.
NOTE
The temperature and pressure values are only given for
information by the Pressio
do not prejudge the health status of the patient.
®
2
CAUTION
®
2
®
The Pressio
2 monitor must only be used with the So-
physa cables, monitoring kits and accessories from the
®
Pressio
2 range.
CAUTION
®
The Pressio
2 monitor must not be used beyond the
period for which its calibration is valid, displayed on the
calibration label on the monitor.
®
io
2 ICP Monitor
®
2 monitoring system. They