Endress+Hauser Prothermo NMT81 Mode D'emploi page 35

Masquer les pouces Voir aussi pour Prothermo NMT81:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Prothermo NMT81
Affichage
Tank level
Navigation
Description
Affichage
Water level
Navigation
Description
Affichage
Courant borne
Navigation
Description
Affichage
Information
supplémentaire
Tension aux bornes 1
Navigation
Description
Affichage
Endress+Hauser
0 ... 1 273,15 K
Application → Valeurs mesurées → Tank level
Shows the actual tank level which is used for average calculation.
–100 ... 100 m
Application → Valeurs mesurées → Water level
Shows the water level which is used for average calulation of product and water
temperature. The value is output as TV via HART. The water level value can be measured
by the water bottom probe if available, or can be sent from the external master. The water
level is measured from the datum plate or tank bottom and includes the end of probe to
zero distance.
–100 ... 100 m
Application → Valeurs mesurées → Courant borne
Affiche la valeur actelle de la sortie courant en cours de mesure.
0 ... 30 mA
Le courant de borne du NMT81 est fixé à 4 mA. D' a utres valeurs peuvent apparaître
pendant la mise à jour du firmware. Si une valeur différente de 4 mA apparaît à un
quelconque autre moment, ceci peut indiquer la présence d' u ne erreur système ou d' u n
autre défaut.
Application → Valeurs mesurées → Tension bornes 1
Affiche la tension au terminal appliqué à la sortie courant.
0,0 ... 50,0 V
Description des paramètres de l' a ppareil