Distance Water Level Covered; Hysteresis Width; Element Weighting; Element 1 - Endress+Hauser Prothermo NMT81 Mode D'emploi

Masquer les pouces Voir aussi pour Prothermo NMT81:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Description des paramètres de l' a ppareil
Réglage usine

Distance water level covered

Navigation
Description
Entrée
Réglage usine

Hysteresis width

Navigation
Description
Entrée
Réglage usine

Element weighting

Navigation
Description
Sélection
Réglage usine
Element 1 ... 24 weighting
Navigation
Description
20
50 mm
Guide utilisat. → Mise en service → wt level covered
Used this function to define the distance, an element has to be covered by water, before it
is included in average water temperature.
0 ... 1 m
50 mm
Guide utilisat. → Mise en service → Hysteresis width
Defines the hysteresis width for all switching levels, to include or exclude elements in the
averaging algorithm.
0 ... 1 m
10 mm
Guide utilisat. → Mise en service → Element weight
By enabling the element weighting, the average calculation can be adapted to different
tank shapes. When disabled, elements will be weighted same.
• Désactiver
• Activer
Activer
 Guide utilisat. → Mise en service → Elemnt 1 ... 24 weight
Utiliser cette fonction pour régler la pondération de chacun des éléments, selon la forme
de la cuve. Les éléments situés dans un volume de liquide élevé peuvent être pondérés avec
un facteur supérieur à celui des éléments situés dans une faible quantité de liquide.