Cleaning / changing the piston unit
– The piston unit
18
the piston frame by turning it with
the Allen key (Fig. 11).
– The piston cylinder
(Fig. 12) and the entire piston unit
can now easily be removed from the
front (Fig. 13) and cleaned externally
using an alcoholic disinfectant.
Re-lubricate the piston with original
STARLAB piston grease (see section
8 Ordering Information).
– The piston unit should not be
disassembled. In the event of
damage, it should be fully replaced
(see section 8 Ordering Information
for a replacement piston unit).
– Reassemble the pipette in the
reverse order.