7.2 Maintaining Multi-Channel Pipettes

The figures shown below refer to the 100µl model. They may differ slightly for other
pipette models.
Changing the O rings (100µl and 300µl models only)
O-rings for the pipette cone are removed
and replaced using the optional auxiliary
tool (see section 8 Ordering Information).
Disassembling the lower housing of the pipette
– Fully unscrew the lower housing
clockwise (Fig. 9).
– To remove the tip ejector
finger and thumb on either side of the tip
ejector, just below the ejector fixing
will allow you to slightly lift the ejector fixing
which will release the tip ejector
6
– Pull both clips
housing through 180° (Fig. 10).
Fig. 10
16
· ErgoOne · User Manual
squeeze with
5
down and rotate the lower
17
by turning
16
. This
6
.
5
Fig. 9