7.1 Maintaining Single-Channel Pipettes

The figures shown below refer to the 200µl model. They may differ slightly for other
pipette models.
The maintenance work is carried out with the aid of the auxiliary tool supplied (tool
not required for 5ml pipette) (Fig. 3)
Fig. 3
For removing and replacing the seals on the 100µl and 1000µl models
A
For removing and replacing the seals on the 200µl model
B
For loosening the lower housing of pipettes up to the 1000µl model
C
Disassembly and cleaning up to the 1000µl model
– Fully depress the tip ejector button
and pull the tip ejector
(Fig. 4).
– Carefully unscrew the tip holder
by turning the auxiliary tool
wise (ErgoOne
®
be removed without the auxiliary
tool).
2
down
5
7
clock-
12
5ml tip holder can
Fig. 4
ErgoOne · User Manual ·
13