3. Select an option.
Note: For descriptions of all the options, refer to the expanded user manual on the manufacturer's website.
Option
Description
Time
Collects a sample at a set time interval, such as every 15 minutes. Select this
Weighted
option for time-weighted sampling. Time Weighted—Sets The time interval
between samples (1 minute to 999 hours). Take First—Sets the first sample to be
collected immediately or collected after the first time interval.
Flow
Collects a sample at a set flow interval, such as every 100 liters (or 100 gallons).
Weighted
Select this option for flow-weighted sampling. An external flow meter or an optional
flow sensor is necessary. Refer to the sampler documentation to connect a flow
meter or flow sensor.
Flow Source—Selects the source of the flow measurement signal: AUX I/O port
(e.g, AUX-mA), sensor port (e.g., Flow 1) or the optional IO9000 module (e.g., IO
1). A flow measurement supplied through a sensor port or the IO9000 module is
only available for selection if the flow measurement is recorded to the data log.
Refer to
sample when the specified volume has not been measured within the selected
maximum time between samples. Enter the maximum time between samples.
When a flow-based sample is collected, the override timer is set to zero. Target—
Selects the flow volume between samples (counts or analog input signal). Take
First—Sets the first sample to be collected immediately or collected after the first
flow interval.
4.1.5.4 Select the sample volume
Select the volume of a sample. For example, set the sampling program to collect 50 mL samples.
1. Push MENU.
2. Go to Programming>Sample Programming>Sample Volume.
3. Selected Fixed>Volume, then enter the volume of each sample (10 to 10,000 mL).
Note: For a description all the options, refer to the expanded user manual on the manufacturer's website.
4.1.5.5 Select the sample distribution—multiple bottles
Select the sample distribution among the bottles.
1. Push MENU.
2. Go to Programming>Sample Programming>Distribution.
Configure data logging
on page 11. Override—When enabled, collects a
English 15