K-MINC-1000 Instructions for Use
Guidance and manufacturer's declaration – electromagnetic emissions
The MINC Mini Incubator is intended for use in the electromagnetic environment specified below. The customer or the end user of the MINC Mini Incubator
should assure that it is used in such an environment.
1
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage
fluctuations/flicker
Complies
emissions
IEC 61000-3-3
Proximity fields from RF wireless communication equipment
Test frequency
Band
a
(MHz)
(MHz)
385
360-390
450
430-470
710
745
704-787
780
810
870
800-960
930
1720
1845
1700-1990
1970
2450
2400-2570
5240
5500
5100-5800
5785
Note: if necessary to achieve the IMMUNITY test level, the distance between the transiting antenna and the ME equipment or ME system may be reduced to
1m. The 1m test distance is permitted by IEC 61000-4-3.
For some services, only the uplink frequencies are included.
a
The carrier shall be modulated using a 50% duty cycle square wave signal.
b
As an alternative to FM modulation, 50% pulse modulation at 18Hz may be used because while it does not represent actual modulation, it would be worst case.
c
Electromagnetic Environment Guidance
The MINC Mini Incubator uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The MINC Mini Incubator is suitable for use in all establishments, including domestic establishments and
those directly connected to the public low voltage power supply network that supplies buildings used for
domestic purposes.
Service
a
Modulation
b
Pulse
TETRA 400
modulation
b
18 Hz
FM
C
GMRS 460
±5 kHz deviation
FRS 460
1 kHz sine
Pulse
LTE Band 13,17
modulation
b
217 Hz
GSM 800/900
Pulse
TETRA 800
modulation
b
iDEN 820
217 Hz
CDMA 850
LTE Band 5
GSM 1800
Pulse
CDMA 1900
modulation
DECT
b
217 Hz
LTE Band 1, 3,
4, 25; UMTS
Bluetooth
Pulse
WLAN
modulation
b
802.11 b/g/n
217 Hz
RFID 2450
LTE Band 7
Pulse
WLAN 802.11
modulation
b
a/n
217 Hz
1-25
Maximum
Distance
Immunity test
power
(m)
(W)
1.8
0.3
2
0.3
0.2
0.3
0.3
2
0.3
2
0.3
2
0.3
0.2
English
Minimum
level
separation distance
(V/m)
(m)
27
0.3
28
0.3
9
0.3
0.3
28
0.3
28
0.3
28
0.3
9