Calibration; Error Code - Baxtran STD Manuel D'usager

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

User's Guide
1. Turn off the power if you are using the scale.
2. Then turn on the power, press
3. Display shows " Pn--- ", waiting for input password.
4. Input password
5. Display will shows " FO d iu ", press
6. Press
7. Empty the hook, after STABLE indicator on, press
completely.
8. After zero calibrate, display shows span calibrate mass weight, you can change this weight, (use
key to move the active digit, use
After input calibrate mass weight, press
on hook, after stable indicator on, press
completely.
ERROR
CODES
- - oL -
Err 4
Err 5
Err 6
Err 10
Err 11
Lo bA
This scale is warranted against defects of manufacturing and materials for a period of 1 year, from the
delivery date.
During this period, GIROPÈS SL will take charge of repairing the scale.
This warranty does not cover defects or damaged caused by misuse, overloading or improper installa-
tion contrary to the recommendations described in this manual.
This warranty does not cover shipping costs for the reparation of the balance.
26
key;
key and
key to enter calibrate, display shows " ULoAd ".
DESCRIPTION
Over range.
Zero Setting Error.
Keyboard Error.
A/D out of range.
No single has been receive-
dl.
The protocol is not right or
suitable.
Battery low.
key at the same time during selfchecking.
key. then press the
key until display shows " F8 CAL ".
key to sure, after some seconds, zero calibrate
key to increment a digit).
key to sure, display show " LoAd ", add calibrate mass
key again, please wait some seconds, span calibrate
RESOLUTION
Remove weight from the scale.
If the problem persist contact your dealer or Taiwan scale for as-
sistance.
The scale was outside the normal auto zero setting range (4% of
capacity), please check the hook is empty, or calibrate zero point
again.
Keyboard short, please check the key wire or whether the key
have mechanical problem.
The values from the A/D converter are outside the normal range.
Remove weight from the scale if overloaded, make sure the pan
is attached.
Indicates the load cell or the electronics may be faulty.
If the problem persist contact your dealer or Taiwan scale for as-
sistance.
Check if the distance is too long or the environment is too bad.
Update the new protocol.
Charge battery immediately.

10. CALIBRATION

key. then press the.
11. ERROR CODES