11. CALIBRATION
GO TO THE SUBMENU OF CALIBRATION, INSIDE THE MENU OF CONFIGURATION.
When the message "F1 CAL" appears in the menu of parameters, press the key
The display is going to show the message "unLoAd".
Remove all the weight from the plate.
Press the key
Introduce the weight of calibration that you want to use by using the keys "0,1,2,...,9" and then put it on the plate. The indicator of stability is going to
switch on to show that the weight is steady.
Next, press the key
Next, the display of weight is going to show the message "LoAd", place on the platform a sample of weight equal to the calibration weight that has been
previously selected and press the key
When the calibration of the scale finishes, it proceeds to do a self-verification test to check if the process has been done correctly. If the display shows
the message "Err 4" you must proceed to repeat the calibration process.
28
This key sets up the auto zero margin. Press the key
(0,5d, 1d, 2d, 4d). Press the key
Establish the new password number. The display is going to show the message "Pin 1". Introduce
the new password number and press the key
Introduce the password again and press the key
message"donE" to show that the new password has been accepted. Keep the new number of the
password in a safe place.
to do the zero adjustment.
USER'S GUIDE BC-BCP
to browse through the different options
to accept the
selection.
. The display is going to change to "Pin 2".
again. The display is going to show the
.
EN