W W e e i i g g h h i i n n g g a a p p a a t t i i e e n n t t
C C A A U U T T I I O O N N
• Always make sure external equipment is removed from product litter when zeroing the scale or weighing the patient.
• Zero the scale at least once a week to maintain scale accuracy.
To weigh a patient:
1. Press and release the W W e e i i g g h h / / o o n n button (E) (see Scale control panel (option) (page 15)) to turn the scale on. Three
dashes will flash on the display for about 10 seconds followed by a weight reading.
2. Press and hold the Z Z e e r r o o button (C) (see Scale control panel (option) (page 15)) for about two seconds.
N N o o t t e e - - Patient must be moved from product to zero the scale.
3. Release the Z Z e e r r o o button and wait for a zero weight to display.
4. Place the patient on the product and press and release the W W e e i i g g h h / / o o n n button. Three dashes will flash on the display for
about 10 seconds followed by the patient weight.
N N o o t t e e - - The
icon may display when using the scale and indicates the product is unstable and unable to weigh.
Stabilize the product and repeat the process.
EN
16
B
D E
Displays patient weight or error messages
Changes the unit of measurement
Sets the scale to zero and erases the previous memory
Add or remove equipment without affecting patient weight
Activates the scale
Store the most recent patient weighing or display the stored weight
C
F