8-1 Preparation
1) Wear a eye protection, a mask and gloves to avoid infection.
2) Remove the bur.
3) Disconnect the handpiece from the motor.
8-2 Cleaning
Manual Cleaning (External Surface)
1) Remove dirt and debris on the handpiece surface using a brush under running water (at 38°C or less).
Do not use a wire brush.
2) Wipe off the moisture on the surface using absorbent cloth.
CAUTION
• If water comes inside the handpiece, blow air into the rear of the handpiece using an air syringe and remove the remaining water.
14