General product specification information
NOTICE
highlighted to avoid product malfunction and
performance reduction.
• Clean and lubricate the handpiece immediately
(within 1 hour) after each treatment to remove
residue. Failure to properly maintain the
handpiece may cause infection, product failure,
overheating leading to burn injuries. (Refer to "8
Post-use Maintenance")
• If blood infiltrates inside a handpiece, clean and
lubricate the handpiece immediately using the
PANA SPRAY Plus or clean the handpiece using
a thermo-disinfector.
2
WARNING
• Depressing the Push Button while the handpiece
is in rotation may lead to overheating, causing
burn injuries or product failure. Avoid the push
button to contact with any oral tissue.
• Always operate with coolant water. Without
coolant water, the handpiece may result in
overheating which could cause a burn injury.
• Use the motor that supplies coolant air.
• Do not use the handpiece for polishing purposes.
Polishing paste could enter the handpiece,
causing the failure of the Push Button or burn
injuries due to the overheating of the handpiece
head.
WARNING