Water Quality Control Unit - LU-VE Emeritus EHVD Serie Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Water quality control unit

Note: the device for measuring the pH is usually set at the factory and is ready for use. If the error message TEMP appears then
a parameter must be modified by following these instructions:
press and hold the ℗ key for at least 2 seconds
scroll using the UP/DOWN arrows to reach the ADMIN level
press the ℗ key
insert code 20
press the ℗ key
scroll using the UP/DOWN arrows to reach the USER level and the sub-level TEMP (confirm the moves with the ℗ key)
scroll using the UP/DOWN arrows to reach the SENS.T parameter
press the key ℗ to change the value
change the value from 1 to 0 (using the UP/DOWN arrows)
press the ℗ key to confirm the value
press the EXIT key to return to the start screen.
This system includes a control unit for the quality of the water,
consisting of a pH probe and an electrical conductivity probe.
The continuously measured values are recorded by the control
system.
When the measured values exceed the warning threshold, a message
is sent by serial communication and displayed on the control screen.
When the values exceed the alarm threshold, apart from sending and
displaying messages as before, the control unit blocks the supply of
water to the system.
During winter shutdown (ambient temperature <5°C) the control
unit has to be put into safety standby condition.
Even after the rest of the unit has been completely drained (ramps,
water distribution piping etc.), some liquid remains (deliberately) in
the probes reservoir of the control unit. This is in order to keep the
probes immersed in liquid (necessary for their conservation,
especially the pH probe).
To avoid freezing causing damage to the system, the cap on the top
of the reservoir must be removed and ethylene glycol added until
the "full" level is reached. Replace the cap and wrap the reservoir
with protective light-obscuring material.
Warning: the probes must always be immersed in liquid (water or
antifreeze mixture).
The default set temperature has little effect on the measured
pH value, within the range used by the system.
10

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Emeritus ehld serieEmeritus xxld serie

Table des Matières