Chapter 5, Error Messages - MSA Galaxy Manuel D'instructions

Système automatisé de tests
Masquer les pouces Voir aussi pour Galaxy:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Chapter 5,
Error Messages
Table 5-1. Error Messages
ERROR MESSAGE
IR FAIL
ABORTED TEST
INST REMOVED
PUMP FAULT or
FLOW FAULT
CAL FAILED or
ZERO FAILED
PARTIAL TEST
DATABASE FULL
DUPLICATE SN
INST SW VER
NON-MASTER
OCCURS WHEN:
Test System hardware failure of IR communications
User aborted test via push-button
Instrument removed from Test System
Pump blocked during test, empty cylinders, incorrect
setup of solenoids, or internal flow error
Instrument sensor cannot be calibrated, bad sensor,
incorrect setup of solenoids, or bad cylinder
Test System lacks gases for full test
Cannot add instrument to database (100 already entered)
Instrument already exists with the same serial number
Instrument needs newer firmware for use on Test System
When using a network interface or shared memory card,
some options/menus are only accessible from the
"rightmost" or master Test Stand
5-1

Publicité

Table des Matières
loading

Table des Matières