40 MHz Analog Oscilloscope HM400
All data valid at 23 °C after 30 minutes warm-up.
Vertical Deflection
Operating Modes:
Channel 1 or 2 only
Channels 1 and 2 (alternate or chopped)
Sum or Difference of CH 1 and CH 2
Invert:
CH 2
XY Mode:
CH 1 (X) and CH 2 (Y)
Bandwidth (-3 dB):
0...40 MHz
DC, 5 mV/div....20 V/div.:
AC, 5 mV/div....20 V/div.:
2 Hz...40 MHz
DC, 1...2 mV/div.:
0...10 MHz
2 Hz...10 MHz
AC, 1...2 mV/div.:
Rise Time (calculated):
<35 ns (1...2 mV/div.)
<8.75 ns (5 mV/div....20 V/div.)
Deflection Coefficient:
1–2–5 Sequence
±5 % (1...2 mV/div.)
±3 % (5 mV/div....20 V/div.)
Variable (uncalibrated):
>2.5:1 to >50 V/div.
Input Impedance:
1 MΩ II 15 pF
Input Coupling:
DC, AC, GND (ground)
Max. Input Voltage:
400 V (DC + peak AC)
Triggering
Automatic:
Linking of peak detection and trigger level
Min. signal height
0.5 div.
Frequency range
5 Hz...50 MHz
Level control range
From peak- to peak+
Normal (without peak):
Min. signal height
0.5 div.
Frequency range
0...50 MHz
Level control range
-10....+10 div.
Slope:
Rising or falling
Sources:
Channel 1 or 2, Line and External
Coupling:
AC (5 Hz...80 MHz),
DC (0...80 MHz),
LF (0...1.5 kHz)
Trigger Indicator:
LED
External Trigger:
Input Impedance:
1 MΩ II 15 pF
External Trigger Signal: 0.3 V
DC (0...50 MHz),
AC (20 Hz...50 MHz)
Max. input voltage:
100 V (DC + peak AC)
Active TV sync. separator:
Field and Line, +/-
Horizontal Deflection
Time Base:
100 ns/div.....0.2 s/div (1–2–5 Sequence)
Accuracy:
±3 %
Variable (uncalibrated):
>2.5:1 to >1.25 s/div.
X Magnification x10:
up to 10 ns/div.
Accuracy:
±5 %
Hold-Off Time:
variable to approx. 10:1
XY
Bandwidth X amplifier:
0...2.5 MHz (-3 dB)
XY Phase shift <3°:
<120 kHz
Operation/Readout/Control
Manual:
via controls and buttons
Autoset:
automatic signal related parameter settings
Save and Recall:
6 instrument parameter settings
Component Tester
Specifications
Test Voltage:
approx. 7 V
Test Current:
max. 7 mA
Test Frequency:
approx. 50 Hz
Test Connection:
2 banana jacks 4 mm Ø
One test circuit lead is grounded
via protective earth (PE)
Miscellaneous
CRT:
D14-363GY, 8 x 10 div. with internal graticule
Acceleration Voltage:
approx. 2 kV
Trace Rotation:
adjustable on front panel
Z-Input (Intens. modulation): max. +5 V (TTL), 10 kHz
Probe ADJ Output:
1 kHz/1 MHz Square Wave Signal approx.
0.2 V
Power Supply (Mains):
105...253 V, 50...60 Hz ±10 %, CAT II
Power Consumption:
approx. 30 W at 230 V/50 Hz
Safety class:
Safety class I (EN61010-1)
HM400E/180112 · C&E · Subject to change without notice · © HAMEG Instruments GmbH
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · info@hameg.com
≤5 V,
pp
(open circuit)
rms
(short-circuit)
rms
(tr <5 ns) for probe adjustment
pp
Operating temperature:
+5...+40 °C
Storage temperature:
-20...+70 °C
Rel. humidity:
5...80 % (non condensing)
Dimensions (W x H x D):
285 x 125 x 380 mm
Weight:
approx. 4.8 kg
Accessories supplied: Line Cord, Operating Manual, 2 Probes 1:1/10:1 (HZ154)
with LF/HF adjustment, CD
+5 °C...+40 ° C
Recommended accessories:
HZ20
Adapter, BNC to 4 mm banana
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1 m
HZ45
19"-Rackmount Kit 4RU
HZ51
Probe 10:1 (150 MHz)
HZ52
Probe 10:1 RF (250 MHz)
HZ53
Probe 100:1 (100 MHz)
HZ100
Differential probe 20:1/200:1
HZ109
Differential probe 1:1/10:1
HZ115
Differential probe 100:1/1000:1
HZ200
Probe 10:1 with auto attenuation ID (250 MHz)
HZ350
Probe 10:1 with automatic identification (350 MHz)
HZ355
Slimline probe 10:1 with automatic identification (500 MHz)
HZO20
High voltage probe 1000:1 (400 MHz, 1000 V
HZO30
Active probe 1 GHz (0.9 pF, 1 MΩ, including many accessories)
HZO50
AC/DC Current probe 20 A, DC...100 kHz
HZO51
AC/DC Current probe 1000 A, DC...20 kHz
®
· DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
S p e c i f i c a t i o n s
)
rms
Subject to change without notice
27