TRIGger[:SEQuence
{[1]|2|3|4}]:HYSTeresis
[:STATe]
TRIGger[:SEQuence[1]|3|4]
: HOLDoff
TRIGger[:SEQuence
{[1]|2|3|4|}]:LEVel
TRIGger[:SEQuence
{[1]|2|3|4}]:RUN:STATe
TRIGger[:SEQuence
{[1]|2|3|4}]:SLOPe
(Command/Query)
The TRIG:HYST <hysteresis> command sets the amplitude of the hysteresis which rejects
the noise associated to the trigger main source.
<hysteresis> is a value at NR1 format taking following values :
0: no noise reject, hysteresis is about 0.5 div.
3: activated noise reject, hysteresis is about 3 div.
To the question TRIG:HYST?, the instrument returns the amplitude of the hysteresis which
rejects the noise associated to the trigger main source.
(Command/Query)
The TRIG:HOLD <time|MAX|MIN|UP|DOWN> command sets the inhibition time of the trigger
(Holdoff).
<time> is a value in format <NRf>, it may be followed or not by a multiple and by the unit.
By default the value is expressed in second.
To the question TRIG:HOLD?, the instrument returns the trigger Holdoff time.
Response format:
<measured value><NL>
value in format <NR3> expressed in second.
(Command/Query)
Used in the Seq. 1 to 4, the TRIG:LEV <level|MAX|MIN|UP|DOWN> command sets the trigger
level of the main source.
<level> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in volt.
To the question TRIG:LEV?, the instrument returns the trigger level of the main source in
SEQuence1.
Response format:
<measured value><NL>
value in format <NR3> expressed in volt.
(Command/Query)
The TRIG:RUN:STAT <1|0|ON|OFF> command starts or stops the acquisition.
ON|1
acquisition starts.
OFF|0
acquisition is stopped.
To the question TRIG:RUN:STAT?, the instrument returns the trigger status.
(Command/Query)
TRIG:SEQ{[1]|2|3|4}:SLOP <POSitive|NEGative> determines :
in SEQuence2 : determines the polarity of the pulse
 POSitive: positive pulse
 NEGative: negative pulse
To the question TRIG:SEQ{[1]|2|3|4<}:SLOP?, the instrument returns the polarity trigger front
or pulse according to the selected SEQuence.
In the other sequences: used to measure the triggering edge of the main source:
 POSitive: rising front
 NEGative: falling front
106
Programmation à distance