RS232 Interface

Dental workstation → PC
An extended Mouse System Protocol is used to transmit defined conditions or commands from the dental workstation to a connected PC.
For the camera and foot switch conditions in byte 1, the bits 0-2 are used to control the video application, to transmit additional commands,
the bits 3-6 are used. The bytes 2-5 are 0 and are not used.
7
6
5
4
3
2
1
User command
Video function
Byte 1 of the Mouse System
In each case only one user command or information to the video function can be transmitted in one data package. The bits of the other
block are set to 0.
The following events and interface commands are defined:
The commands are generated in each case when the event is triggered.
Rhomb key activated
Rhomb deactivated
Call key
X-ray image viewer key
CCW rotation key
(with instruments deposited or video camera removed)
Instrument light key
(with instruments deposited or video camera removed)
Quick setting key 100
(with instruments deposited or video camera removed)
Quick setting key 75
(with instruments deposited or video camera removed)
Quick setting key 50
(with instruments deposited or video camera removed)
Quick setting key 25
(with instruments deposited or video camera removed)
Quick setting key 1
(with instruments deposited or video camera removed)
Forward cursor from chair key block
Forward cursor from instrument key block
Spray key
Left discrete C+ FS button active with camera removed
Chip blower key
Right discrete C+ FS button active with camera removed
Camera removed
Camera deposited
Foot switch active with camera removed
Foot switch inactive with camera removed
When video camera is assembled on HE side and when in program SIUCOM.exe mode: "Sirocam C direct" is selected:
Press foot switch
Press button on camera
PC → Dental workstation
The RS232 configuration and service information can be output via PC with an 1-byte command.
Command
0DH
Data transmission rate:9,600 Baud
59 02 981
1
0
Event
Action
Output of configuration and
service information
Blatt
12A
Sheet
Feuille
Hoja
Remark
only if switching mode
also be operated via FS
cannot be operated via FS
cannot be operated via FS
cannot be operated via FS
also be operated via FS
only be operated via FS
only be operated via FS
Remark
RS232 command 32
(1st byte of the 5-
byte protocol)
88H
90H
98H
A0H
A8H
B0H
B8H
C0H
C8H
D0H
D8H
E0H
E8H
F0H
F8H
81H