INSTALLATION
"alYa" Dental light, version S/tS
Check that the package contains the following components:
- Dental light (in the version requested)
- Lever envelope + tool
- Instruction manual
lamp assembly, combined version
- Install the lamp by inserting the terminal lamp pin into the specific hole on the combined unit.
Arm length 855 mm
Arm length 550 mm
1) PoWEr CaBlE:
- lIGHt WItH tranSForMEr; must be powered directly from the mains, which
should correspond to the specifications indicated on the data plate or in the technical spe-
cifications of the manual.
- Ensure that the line switch conforms to the surface distances in air as specified in Interna-
tional Publication IEC 328.
- lIGHt WItHout tranSForMEr; must be powered with low-voltage alternating
current (24V AC) using a safety transformer that conforms to EN 60601-1.
2) rEMotE Control CaBlE
- Connect the cable with two push-buttons (not supplied), whose contact is normally open,
according to the following scheme:
16
CONNECTIONS
ANd
the system must be cleaned before use.
(see the paragraph Cleaning the device)
the system must be installed by specialized technicians.
the power must always be off when the system is installed.
Simultaneous use of the lamp with an electric scalpel can cause the scalpel
to malfunction. to reduce risks, it is advisable to use the version controlled
with the joystick instead of the proximity sensor. always verify compatibility
before use.
progeCAD 2008 Smart! - Private use only
SAFE WORKING LOAD MINIMUM BREAKING LOAD
3.12 Kg
2.64 Kg
(where available):
26 Kg.
35 Kg