Information On Mounting And Operation In Hazardous Areas - WIKA TR10-0 Mode D'emploi

Masquer les pouces Voir aussi pour TR10-0:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7. Information on mounting and operation in hazardous areas

7. Information on mounting and operation in hazardous areas
WARNING!
Non-observance of these instructions and their contents may result in the loss of
explosion protection.
The requirements of IECEx scheme and ATEX directive must be followed.
Additionally the specifications of the respective national regulations concerning
Ex usage (e.g. IEC/EN 60079-10 and IEC/EN 60079-14) apply.
The responsibility for classification of zones lies with the plant operator and not the
manufacturer/supplier of the equipment.
The plant operator guarantees, and is solely responsible, that all thermometers in use
are identifiable with respect to all safety-relevant characteristics. Damaged thermometers
may not be used.
For the installation of the thermometer, only components (e.g. cables, cable glands, etc.)
permitted for "flameproof" may be used.
For earthing the conductive screen, follow the specifications of IEC/EN 60079-14.
When using a transmitter/digital display, note and follow:
- The contents of these operating instructions and those of the transmitter/digital display
- The relevant regulations for installation and use of electrical systems
- The regulation and guidelines regarding explosion protection
The flameproof thermometers should only be fitted to certified housing- or connection
heads certified with a "flameproof" ignition protection type.
For fitting, the permitted flameproof joints for electrical equipment for gas hazardous
areas are contained in IEC/EN 60079-1. Flameproof joints
be ≥ 5 mm for housing volumes < 100 cm³ and ≥ 8 mm for housing volumes > 100 cm³.
There must be ≥ 5 threads engaged.
Flameprooof joints
There must be ≥ 3.5 threads engaged. These specifications for flameproof joints must be
adhered to, without fail, when fitting and during operation.
The direct threaded connection of the thermometer to the connection head or housing
must not be twisted or opened. Any alignment of the housing may only be made using the
optional "nipple-union-nipple" neck tube.
The temperature resistance of the connecting cable must match the permissible
operating temperature of the housings.
For ambient temperatures above 60 °C, heat-resistant connecting cable must be used.
No batteries may be built in to the flameproof housing.
1) Section 5.3 of IEC 60079-1
2) In accordance with table 3 of IEC 60079-1
WIKA operating instructions models TR10-0, TC10-0 (Ex d)
for tapered threads
1)
for parallel threads
1)
, must have ≥ 5 available threads on each part.
2)
EN
, must
2)
15

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Tc10-0

Table des Matières