6.2 Darkfield obervation (DF)
Usable objectives with different inserts
Insert
DF
Top lens position
DF OIL
Top lens position
(*) For observation with a 100X objective in darkfield oil we recommend the use of a 100X oil objective
with iris.
Remove all filters from the optical path.
1. Rotate the condenser turret to enter the "DF"
position.
2. Pull the lever of the polarizer ① to remove it from
the optical path. (Fig. 11)
3. Place a specimen on the stage and focus the
specimen.
4. Observing in the eyepieces, lower or raise the
condenser until a homogeneous illumination of
the sample is obtained and therefore an optimal
dark field effect is achieved.
5. Using the centering screws on the condenser ①
(Fig. 13) centre the dark field ring until the field of
view is evenly illuminated.
Darkfield requires a huge amount of light.
Switching from darkfield to brightfield, one
could be dazzled. Do not keep your eyes on
the eyepieces when moving the condenser
turret from DF to BF.
"Dry" darkfield observation, that is, without
the use of oil, is only possible with objectives
with N.A. lower than 0,7.
Observing in darkfield, it may be necessary to
slightly raise the condenser from the normal
position to obtain a more homogeneous
illumination. This is not a defect.
Objective
4X
-
OUT
-
OUT
10X
20X
¡
¡
IN
-
-
-
Page 11
40X
60X
-
¡
-
-
¡
IN
100X
-
-
¡(*)
F ig. 13
F
ig. 13

Publicité