13. The unit is noisy
Ball screw is noisy
grease the ball screw every 1000 hours
readjust the ball screw with the motor
change the ball screw
Mechanical articulations are noisy
grease mechanical articulations
14. Offset between mechanical angle and value of potentiometers
check if the potentiometer's knob is correctly fitted
verify if the copy potentiometer and the cotter pin are tightened
redo the adjustment of the unit (see method "MET09057" for more details)
adjustment potentiometers failure: change the motor PCB
- Stabilized power supply (0-30V DC / 2.5A)
- Amperemeter, ohmmeter, voltmeter
- Allen wrenches
- Pin drift (2.5mm)
- Flat screwdrivers
- Grease
Ball screw: MOBILUX EP2
Sliding parts on the nut support: MOLYKOTE M55
Rubber strips: MOLYKOTE PG54
Knee articulation points: MOLYKOTE PG75
Please do not include pads when returning the KINETEC PRIMA ADVANCE to our workshops due to health
and safety standards within our company. These pads will be systematically incinerated.
If you want repair information you must have the serial number of the unit.
REQUIRED TOOLS
RETURNING THE UNIT TO OUR WORKSHOPS
REPAIR INFORMATIONS
- Cruciform screwdrivers
- Goniometer (angle adjustment)
- 0 - 60 daN dynamometer (load adjustments)
- Dynamometer attachment tools.
- Silicone: LOCTITE 5900 (black)