PAR. 10
CYCLES
Fragile unwrapped solids
121°C
SOLIDS
Fragile wrapped solids
Fragile hollow
121°C
instruments, unwrapped
HOLLOW
Fragile hollow
instruments, wrapped
Unwrapped solids
134°C
SOLIDS
Wrapped solids
Unwrapped hollow
134°C
instruments
HOLLOW
Wrapped hollow
instruments
Wrapped/unwrapped
134°C
Solid and hollow material
PRION
121°C
Porous material
POROUS
134°C
Unwrapped
FLASH
Solid/hollow
BOWIE& DICK
Material
HELIX
None
VACUUM
TEST
VARIABLE
N.B.: The drying time can be extended by 5 minutes for each sterilisation cycle. To do this, start the
cycle by pressing START/STOP button and then press SELECT button for 10 seconds until the
acoustic signal. The total cycle time shown on the Display will have increased by 5 minutes. Always
set extra drying time when you run the autoclave at full wrapped load.
It is obligatory to wrap invasive instruments (e.g. surgical instruments) to ensure higher sterility
levels. All the data in this table are indicative; the choice of the sterilisation cycle must be based on
the data supplied by the manufacturer of the object to be sterilised.
STERILISATION TABLE
MATERIALS
STERILIZATION
3 min. 30 sec.
DRYING
TIME
TIME
18 minutes
10 minutes
18 minutes
10 minutes
5 minutes
10 minutes
5 minutes
10 minutes
20 minutes
10 minutes
15 minutes
18 minutes
4 minutes
15 minutes
39
MAX. LOADING.
WORKING
PRESSURE
6 Kg
(1.12÷1.14)
unwrapped load
3 Kg
wrapped load
6 Kg
(1.12÷1.14)
unwrapped load
3Kg
wrapped load
6 Kg
(2.12÷2.14)
unwrapped load
3 Kg
wrapped load
6Kg
(2.12÷2.14)
unwrapped load
3 Kg
wrapped load
6 Kg
(2.12÷2.14)
unwrapped load
3 Kg
wrapped load
4 Kg
(1.12÷1.14)
Total full load
6 Kg
(2.12÷2.14)
empty
empty
empty
bar
bar
bar
bar
bar
bar
bar
(-0.86)
bar