Each grain can be individually adjusted by 10.0% in increments of 0.1% to more
closely match the results of an elevator tester. The calibration adjustment typi-
cally compensates for the differences in the variety/hybrid of grain you are testing
relative to the varieties used to develop the grain moisture calibration in the tester.
TO ADJUST GRAIN OFFSET:
1. Turn on Unit (press ON/OFF button)
2. Use UP/DOWN buttons to scroll to "GRAIN OFFSET" then press ENTER
3. Use UP/DOWN buttons to make the desired adjustment
4. Press ENTER to save this offset (or press back arrow to not save it ).
You are now back to active grain display where you are ready to test
Grain Offset
NOTE: The grain offset adjustment (bias/calibration) is applied to the last grain
you tested, since a test must be made prior to knowing what adjustment to
apply.

CLEAR GRAIN OFFSET

1. Turn on Unit (press ON/OFF button)
2. Use UP/DOWN buttons to scroll to "GRAIN OFFSET" then press ENTER
3. Use UP/DOWN buttons to make adjustment 0.0
4. Press ENTER (after displaying the save [disc] icon the unit takes you back
to the active grain display and ready to test)
Operation
Figure 10. GRAIN OFFSET
Figure 11. GRAIN OFFSET
12