6. Packaging / 7. Starting, operation

6. Packaging

Has everything been supplied?
Check the scope of supply:
Completely assembled pressure transmitters.
With flush version (S-11) including pre-assembled sealings and protection cap.
Inspect the pressure transmitter for possible damage during transportation. Should
there be any obvious damage, inform the transport company and WIKA without delay.
Keep the packaging, as it offers optimal protection during transportation (e.g. chan-
ging installation location, shipment for repair).
Ensure that the pressure connection thread and the connection contacts will not be
damaged.
In order to protect the diaphragm, the pressure connection of the instrument S-11 is provided
with a special protection cap.
Remove this protection cap by hand only just before installing the pressure transmitter
in order to prevent any damage to the diaphragm or the thread.
Keep the protection cap of the pressure connection thread and the diaphragm for
later storage or transport.
Mount the protection cap when removing and transporting the instrument.

7. Starting, operation

Required tools: wrench (flats 27), screw driver
Diaphragm test for your safety
It is necessary that before starting the pressure transmitter you test the diaphragm visually, as
this is a safety-relevant component.
Pay attention to any liquid leaking out, for this points to a diaphragm
!
damage.
Check the diaphragm visually for any damage (S-11).
Warning
Use the pressure transmitter only if the diaphragm is undamaged.
Use the pressure transmitter only if it is in a faultless condition as far as the
safety-relevant features are concerned.
6 WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio S-10, S-11
GB
7. Starting, operation

Mechanical connection

max.
50 Nm
Sealing
Generally the serial number on the product label applies. If there is no serial number on the product label,
the number on the hexagon will apply.
Remove the protection cap by hand only just before installation and absolutely avoid
any damage to the diaphragm during installation as well (S-11).
For Model S-10 you have to provide for a sealing element; exceptions are instruments
with self-sealing threads (e.g. NPT thread). For Model S-11 the sealing is included in
delivery.
Please refer to our data sheet "Pressure gauge sealing washers AC 09.08" in WIKA's
product catalog Pressure and Temperature Measurement or our website
www.wika.de for details about sealing washers.
When mounting the instrument, ensure that the sealing faces of the instrument and
the measuring point are clean and undamaged.
Screw in or unscrew the instrument only via the flats using a suitable tool and the
prescribed torque. The appropriate torque depends on the dimension of the pressure
connection and on the sealing element used (form/material). Do not use the case as
working surface for screwing in or unscrewing the instrument.
When screwing the transmitter in, ensure that the threads are not jammed.
For tapped holes and welding sockets please see Technical Information IN 00.14 for
download at www.wika.de - Download
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio S-10, S-11
Product label (example)
Signal
Power Supply
Serial No.
S #
Product No.
P #
PIN assignment
Coded manufacture date
GB
7