GPIB Interface

The 24-way GPIB connector is located on the instrument rear panel.
The pin connections are as specified in IEEE Std. 488.1-1987 and the instruments in the range
comply with IEEE Std. 488.1-1987 and IEEE Std. 488.2-1987 and contain the following IEEE
488.1 subsets.
GPIB Subsets
This instrument contains the following IEEE 488.1 subsets:
Source Handshake
Acceptor Handshake AH1
Talker
Listener
Service Request
Remote Local
Parallel Poll
Device Clear
Device Trigger
Controller
Electrical
Interface
*
Although no Device Trigger capability is included, the GET message will not cause a
command error unless its position in the input stream dictates that it should; e.g. buried
inside a <PROGRAM MESSAGE UNIT>.
GPIB IEEE Std. 488.2 Error Handling
The IEEE 488.2
If the instrument is addressed to talk and the response formatter is inactive and the input queue
is empty then the
in the Standard Event Status Register, a value of 3 to be placed in the Query Error Register and
the parser to be reset. See the STATUS REPORTING CAPABILITIES section for further
information.
The IEEE 488.2
send a response message and a
or the input queue contains more than one END message then the instrument has been
INTERRUPTED
Standard Event Status Register, a value of 1 to be placed in the Query Error Register and the
response formatter to be reset thus clearing the output queue. The parser will then start parsing
the next
<PROGRAM MESSAGE UNIT>
CAPABILITIES section for further information.
The IEEE 488.2
a response message and the input queue becomes full then the instrument enters the
state and an error is generated. This will cause the Query Error bit to be set in the Standard
Event Status Register, a value of 2 to be placed in the Query Error Register and the response
formatter to be reset thus clearing the output queue. The parser will then start parsing the next
<PROGRAM MESSAGE UNIT>
section for further information.
GPIB Parallel Poll
The power supplies offer complete parallel poll capabilities. The Parallel Poll Enable Register is
set to specify which bits in the Status Byte Register are to be used to form the
The Parallel Poll Enable Register is set by the ∗PRE <nrf> command and read by the ∗PRE?
16
error (addressed to talk with nothing to say) is handled as follows.
UNTERMINATED
error is generated. This will cause the Query Error bit to be set
UNTERMINATED
error is handled as follows. If the response formatter is waiting to
INTERRUPTED
<PROGRAM MESSAGE TERMINATOR>
and an error is generated. This will cause the Query Error bit to be set in the
error is handled as follows. If the response formatter is waiting to send
DEADLOCK
from the input queue. See the STATUS REPORTING CAPABILITIES
SH1
T6
L4
SR1
RL1
PP1
DC1
DT0*
C0
E2
from the input queue. See the STATUS REPORTING
has been read by the parser
DEADLOCK
local message
ist

Publicité