Technical data for the OMRON NE-C801 Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:
Particle Size:
Medication Tank
Capacity:
Appropriate Medication
Quantities:
Sound:
Nebulization Rate:
Aerosol Output:
Aerosol Output Rate:
Result of cascade impactor measurements for particle size**
0.1
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering,
Doshisha University, Kyoto, Japan, according to EN13544-1:2007+A1:2009
Notes:
• Performance may vary with drugs such as suspensions or high viscosity. See
drug supplier's data sheet for further details.
• See website of OMRON HEALTHCARE EUROPE to update technical
information.
URL: www.omron-healthcare.com
**MMAD Approx. 3μm
MMAD = Mass Median Aerodynamic Diameter
7ml maximum
2ml minimum - 7ml maximum
*Approx. 46dB
*Approx. 0.3ml/min (by weight loss)
**0.47ml (2ml, 1%NaF)
**0.06ml/min (2ml, 1%NaF)
Cumulative % particle mass of sodium fluoride under size
1
Particle Size Dp (μm)
Individual
Mean
MMAD 3.0 μm
1
0
9. Technical data
100
90
80
70
60
50
40
30
20
10
0
100
25
EN