NAMUR instructions
Function
IN_NAME
Input description name
IN_PV_X
X=1;2;3;4;5;
Reading the real value
7; 8;
IN_SOFTWARE
Input software ID number, date and version
IN_SP_X
X=1;2;3;4;7;8;
Reading the set rated value
12;42;50;52;53;
54;55;56;
IN_TYPE
Input laboratory unit ID
OUT_NAME name
Output description name. (Max. 6 characters, default: IKAMAG)
OUT_SP_12@n
Setting the WD safety temperature with the echo of the set value
OUT_SP_42@n
Setting the WD safety speed with the echo of the set value
OUT_SP_X n
X=1;2;4;7;50;
Setting the rated value to n
52;53;54;55;56
OUT_WD1@m
Watchdog mode 1: When a WD1 event occurs, the heating and stirring functions are shutdown and message ER 2 is
displayed.
Set the watchdog time to m (20...1200) seconds, with echo of the watchdog time. This instruction starts the watchdog
function and must be sent within the set watchdog time.
OUT_WD2@m
Watchdog mode 2: When a WD2 event occurs, the speed setpoint will be set to the WD safety setpoint speed and the
temperature setpoint will be set to the WD safety setpoint temperature. The WD warning is displayed.
The WD2 event can be reset with OUT_WD2@0 - resetting also blocks the watchdog function. Set the watchdog time
to m (20...1200) secondes, with echo of the watchdog time. This command starts the watchdog function and must be
sent within the set watchdog time.
RESET
Switching off the instrument function.
022009
Display
(additional)
39