Chapter 14
ERROR MESSAGE
Error message will be displayed in the screen if there is something wrong when measuring. The causes and solutions are
shown as follows:
Error Message
Self-test failure
System failure
Loose cuff
Air leakage
Air pressure error
Weak signal
Overpressure
Excessive movement
Over range
Saturated signal
Time out
ENGLISH
Causes
Function abnormal
Cuff is not connected correctly.
Cuff plug fall off
Air pressure error
The pulse signal is too weak or the cuff
is loose.
Cuff is blocked or squeezed
The signal extent is too big owing to the
limbs or body moving or other reasons
when measuring
It takes too much time
Solutions
Please contact us
Correctly connect cuff (refer to Chapter 9)
Make sure the cuff plug is securely inserted
in the windpipe (refer to Chapter 9)
Refer to the troubleshooting
Correctly connect cuff (refer to Chapter 9)
Correctly connect cuff (refer to Chapter 9)
Keep limbs, body still, measure again
68