REF 5500E/5506 series is intended for use in the electromagnetic environment specified below�
The customer or the user of REF 5500E/5506 series should insure that it is used in such an environment�
Immunity test
IEC 60601 test level
3 Vrms
150 kHz to 80 MHz
Conducted RF
6 Vrms
IEC 61000-4-6
150 kHz to 80 MHz (ISM and
amateur radio bands)
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2�7 GHz
27 V/m, 385MHz
28 V/m, 450MHz
Enclosure port immunity IEC
9 V/m, 710 MHz to 780 MHz
61000-4-3
28 V/m, 810 MHz to 2450 MHz
9 V/m, 5240 MHz to 5785 MHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures, objects and people�
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
a
cannot be predicted theoretically with accuracy� To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered� If the
measured field strength in the location in which REF 5500E is used exceeds the applicable RF compliance level above, REF 5500E should be observed to verify normal operation� If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating REF 5500E�
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m�
b
Guidance and manufacturer's declaration — electromagnetic immunity
Compliance level
3 Vrms (150 kHz to 80 MHz)
6 Vrms (150 kHz to 80 MHz)
3 V/m
80 MHz to 2�7 GHz
27 V/m, 385MHz
28 V/m, 450MHz
9 V/m, 710 MHz to 780 MHz
28 V/m, 810 MHz to 2450 MHz
9 V/m, 5240 MHz to 5785 MHz
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer
to any part of REF 5500E, including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter�
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2�7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m)�
Field strength from fixed RF transmitters, as determined by an electromagnetic
site surveya, should be less than the compliance level in each frequency rangeb�
Interference may occur in the vicinity of equipment marked with the following
symbol:
d = 6/E√P
380MHz to 5�8GHz