BD ClipVac Presurgical Hair Removal System Mode D'emploi page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15
Recommended separation distance between portable and mobile RF communications equipment and REF 5500E.
REF 5500E/5506 series is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled� The customer or the user of
REF 5500E/5506 series can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and REF 5500E/5506 series as recommended below, according to the maximum output power of the communications equipment�
Rated maximum output power of
transmitter
W
0�01
0�1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer�
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people�
14
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2 √P
0�12
0�38
1�2
3�8
12
m
80 MHz to 800 MHz
d = 1.2 √P
0�12
0�38
1�2
3�8
12
800MHz to 2.7GHz
d = 2.3 √P
0�23
0�73
2�3
7�3
23

Publicité

Table des Matières
loading

Produits Connexes pour BD ClipVac Presurgical Hair Removal System

Ce manuel est également adapté pour:

5500e5506a5506e5506u5506k

Table des Matières