20. EMC Information
The instrument fulfils the requirements according to EN 60601-1-2 regarding emission and immunity. If sensitive electronic equipment is affected by the
instrument, try to increase the distance to such equipment. The charger should not be connected during measurements.
Guidance and manufacturer's declaration - Electromagnetic Emissions.
Osseo 100 is intended for use in the electromagnetic environment specified below.
Emissions tests
RF emissions C1SPR11
RF emissions C1SPR11
Harmonic emissions IEC61000-3-2
Voltage fluctuations/flicker emissions IEC61000-3-3
Guidance and manufacturer's declaration - Electromagnetic Immunity Test Levels
Osseo100 is intended for use in the electromagnetic environment specified below.
Immunity test
Electrostatic discharge (ESD)
Radiated RF EM fields
Proximity fields form RF wireless communications equipment
Rated power frequency magnetic fields
Electrical fast transient/burst
Surges Line-to-line, Surges Line-to-ground
Conducted disturbances induced by RF fields
Voltage dips, Voltage interruptions and Electrical transient
condition along supply lines
Compliance
Electromagnetic environment - guidance
Group 1
Osseo 100 uses RF energy only for its internal function.
Class B
Osseo100 Rechargeable battery operated device
Not applicable
Not applicable
EMC standard or test method
IEC61000-4-2
IEC61000-4-3
IEC61000-4-3
IEC61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC61000-4-6
IEC 61000-4-11
Professional healthcare facility environment
± 8kV contact
± 2 kV ± 4 kV ± 8 kV ± 15 kV air
3 V/m
80 MHz - 2.7 GHz
80 % AM at 1 kHz
30 cm minimum separation distance from radio transmitter
30 A/m
50 Hz or 60 Hz
± 2kV
100 kHz repetition frequency
± 0.5, ± 1 kV, ± 2 kV
3V
0,15 MHz – 80 MHz
6 V in ISM bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
5% UT, 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0 % UT; 1 cycle
And 70 % UT; 25/30 cycles (50/60Hz)
Single phase: at 0°
0 % UT; 250/300 cycle (50/60 Hz)