USING YOUR SOCLEAN 2
After set-up, using your SoClean 2 is extremely easy.
After following the connection procedure on page 14,
your SoClean 2 maintains your PAP equipment daily.
Figure 6. Install Removable Hose Slot Plug
WARNING: Do not physically alter the Hose Detect
Switches on your SoClean 2. Inspect the Hose Detect
Switches regularly for damage or inability to move up/down.
Failure of the Hose Detect Switch on your SoClean 2 may
result in the SoClean operating at an unexpected time.
1. Open the Lid.
2. Install the Removable Hose Slot Plug into one
open slot (Figure 6).
Orientate the Removable Hose Slot Plug as shown in
Figure 6. The round face of the Removable Hose Slot Plug
is on the outside of the SoClean 2.
Note: Your mask and PAP Hose can insert from either side
of the SoClean 2. To change sides, remove the Removable
Hose Slot Plug by lifting it straight out, and insert it into the
slot on the opposite side.
22 | USING YOUR SOCLEAN 2
3. Remove outer plastic wrapping and blue tape
from the Cartridge Filter and fully seat it in the
Cartridge Filter Slot at the back right corner of the
SoClean Chamber. Ensure proper installation by
orienting the Cartridge Filter vents to face the front of
the machine as pictured in Figure 6.
4. Insert your PAP mask into the SoClean Chamber,
allowing its attached hose to rest in the open Hose Slot,
opposite the Removable Hose Slot Plug (Figure 7).
5. Close the Lid with the hose placed in the open slot
(Figure 8). The SoClean 2 will now automatically operate
according to its preset time and duration.
Note: The SoClean 2 won't operate without the mask
and hose present in the closed Chamber. Additionally, the
SoClean 2 will not operate if the Removable Hose Slot
Plug is missing or improperly seated.
WARNING: Keep the SoClean 2 Lid closed until the green
status light illuminates (approximately 2 hours after completion
of the cycle).
It is necessary to let the mask remain in the closed
SoClean Chamber for at least 2 hours to ensure that
the ozone performs its sanitizing function and dissipates
appropriately. After that period, the status light shines
green and the mask is ready to be removed from the
SoClean Chamber.
Ozone quickly dissipates into the atmosphere. If you
accidentally interrupt the cycle and release ozone by
opening the SoClean Chamber Lid, simply step away
from the SoClean 2 for 5 minutes to avoid unintended
exposure. Note: Your PAP equipment will not be fully
maintained if you interrupt the cycle; therefore, it is
recommended that you restart the cycle.