INDICATIONS FOR USE, CONTRAINDICATIONS,
SPECIFICATIONS, WARNINGS AND CAUTIONS
Indications for use: The SoClean 2 is an Automated
Supplemental PAP Maintenance* Device that is intended
for use by a single-user, for personal at-home-use to
maintain PAP device accessories. SoClean 2 kills 99.9%
of microorganisms* commonly found on PAP masks and
hoses when set at the 12-minute cycle time duration.
SoClean recommends users follow PAP-manufacturer's
instructions for cleaning their PAP accessories. The SoClean
2 is intended for use as supplement to that cleaning.
While the SoClean 2 is connected to your PAP equipment
you can still easily disconnect your mask and hose to
follow your manufacturers daily cleaning instructions.
Contraindications for use: Persons with underlying lung
diseases, such as asthma and chronic obstructive pulmonary
disease (also known as COPD, which includes emphysema
and chronic bronchitis), and those with cardiovascular
disease may be sensitive to ozone and should consult with
their physician before using this product.
Read this User Manual before operating
the equipment.
Use this product only in the manner described in this User
Manual. If the equipment is used in a manner not specified
by the manufacturer, the safety protection provided by the
equipment may be impaired.
*U.S. independent lab testing with SoClean 2 demonstrates 99.9% reduction of the following microorganisms comprised of the following bacteria:
Staphylococcus haemolyticus, Staphylococcus hominis, Staphylococcus epidermidis, and yeast: Candida albicans
** The term "maintenance" herein refers to SoClean 2 performance killing 99.9% of microorganisms* commonly found on PAP masks and hoses when set at
the 12-minute cycle time duration.
2 | INDICATIONS FOR USE, CONTRAINDICATIONS, SPECIFICATIONS, WARNINGS AND CAUTIONS
2 |
Specifications:
Electrical characteristics:
AC adapter input: AC 100~240V, 50/60HZ, 0.5A
AC adapter output: DC 12V, 1.5A max.
Power consumption: 18W max.
Environmental characteristics:
Operating: 10°C to 38°C (50°F to 100°F),
15% to 70% humidity
Storage and transport conditions:
-20°C to +55°C (-4°F to 131°F),
15% to 70% humidity
Ozone concentration:
When operated in a 11.6 m
room with 2.4
2
metre-high ceilings, the average ambient ozone
concentration in the room over the operating
cycle is < 0.05 parts per million (PPM).
Ozone Cycle Run Time Duration
12 minutes for the most effective maintenance
Physical characteristics:
Dimensions: 200 x 184 x 225 mm
Weight: 2.5 kg
Cord length: 142 cm
Data on le.
†
†