7.1. REFERENCE CONDITIONS

Infl uence quantities
Temperature
Relative humidity
Supply voltage
Frequency range
Capacitance in parallel on resistance
Electric fi eld
Magnetic fi eld

7.2. CHARACTERISTICS PER FUNCTION

7.2.1. VOLTAGE
 Characteristics
Measurement range
Frequency range
(4)
Resolution
Accuracy
Input impedance
(4): Above 500 Hz, the small display indicates "- - - -" and the main display gives only a rough indication of the peak
value of the measured voltage.
 Measurement category: 1000 V CAT III or 2500 V CAT I (transients   2.5 kV)
7.2.2. INSULATION RESISTANCE
 Method: Voltage-current measurement per IEC 61557-2
 Nominal output voltage: 500, 1000, 2500, 5000 Vdc (or adjustable from 40 V to 5100 V)
 Adjustment step in variable mode: 10 V from 40 V to 1000 V
 Off-load voltage:   1.02 x Un ± 2% (Un ± 2% in variable mode)
 Nominal current:   1 mAdc
 Short-circuit current: < 1.6 mA ±5%
 Charging current into capacitive component: 3 mAdc approximately when starting measurement
 Maximum acceptable voltage during measurement: Upeak = (1.1 + dISt) Un + 60V

7. SPECIFICATIONS

Reference values
23 ± 3 °C
45 to 55 % RH
9 to 12 V
DC and 15.3 to 65 Hz
< 40 A/m
1.0...99.9 V
100...999 V
DC and 15 ... 500 Hz
0.1 V
1 V
1% +5 pt
750 k  to 3 M  depending on measured voltage
100 V from 1000V to 5100 V
71
0 µF
nil
1000...2500 V
2 V
1% +1pt
with dISt = 0.03 – 0.10 or 0.20
2501...4000 V
DC
2 V