Pressing on the V-TIME key during the measurement alternately displays the measurement time and the
exact voltage generated on the small display (see § 4.2).
The stopping of the measurement is caused by pressing the STOP key. After the measurement stops,
the main result remains displayed.
It is possible to scroll through all the other results available on the main display using the R-DAR-PI-DD
key. This key may also be used before the triggering of the measurement (see § 4.3).
If the "Programmed time test"
results stored automatically (see § 4.2 and 4.3).
If the ALARM function is activated, a buzzer will sound as soon as the measurement exceeds or falls
below the programmed threshold in the SET-UP confi guration menu (see § 4.4).
 Display of values after a measurement
The indications below may be displayed:
R-DAR-PI-DD key
Main display
Resistance
duration (min. sec)
DAR
duration (min. sec)
PI
duration (min. sec)
duration (min. sec)
(1)
DD
duration (min. sec)
Current
(2)
Capacitance
duration (min. sec)
R(t)
(1): The value of DD is displayed only one minute after the measurement stops.
(2): The capacitance measurement (µF) is displayed only after the measurement stops and the circuit
has been discharged.
mode was selected, the R(t) key is used to access all the intermediate
Small display
52
V-TIME key
Small display
if the MR key is activated (C.A 6547)
date, time, test voltage, OBJ:TEST
date, time, test voltage, OBJ:TEST
date, time, test voltage, OBJ:TEST
date, time, test voltage, OBJ:TEST
last test voltage