5.1. REFERENCE CONDITIONS
Influence quantities
Temperature
Relative humidity
Supply voltage
Frequency range
Capacity in parallel on resistor
Electrical field
Magnetic field
5.2. CHARACTERISTICS PER FUNCTION
5.2.1. vOLTAGE
Characteristics
Measurement range
Frequency range
3
Resolution
Accuracy
Input impedance
3: Over 500 Hz, the small display indicates "- - - -" and the main display gives only an assessment of
the peak value of the measured voltage.
5.2.2. CURRENT
Current measurement before the insulation measurement:
Measure-
0.000 -
ment range
0.250 nA
Resolution
1 pA
Accuracy
15% ± 10 pt
5. SPECIFICATIONS
1.0 - 99.9 V
DC and 15 Hz - 500 Hz
0.1 V
1% ± 5 pt
750 kΩ at 3 MΩ depending on measured voltage
0.250 -
10.00 -
9.999 nA
99.99 nA
1 pA
10 pA
10%
Reference values
23 ± 3°C
45 to 55% RH
9 to 12 V
DC and 15.3...65 Hz
0 µF
nil
< 40 A/m
100 - 999 V
1000 - 2500 V
15 Hz - 500 Hz
1 V
100.0 -
1000 -
999.9 nA
9.999 µA
100 pA
1 nA
5%
1000 - 5100 V
2 V
1% ± 1 pt
10.00 -
100.0 -
99.99 µA
999.9 µA
10 nA
100 nA
DC
2 V
1000 -
3000 µA
1 µA
10%