Use; Measurement Procedure; Insulation Measurement - Chauvin Arnoux C.A 6545 Notice De Fonctionnement

Masquer les pouces Voir aussi pour C.A 6545:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1

5.1. MEASUREMENT PROCEDURE

 Start up the instrument by setting the switch to the desired measurement position. All the segments
of the LCD screen are displayed, then the battery voltage.
 Connect the + and - leads to the measurement points.
 The input voltage is measured continuously and indicated on the small display.
If an external voltage higher than the authorised limit value (see § 3.2) is present, the
measurement will be inhibited.
 A press on START/STOP starts the measurement.
 Another press on START/STOP stops the measurement. The last result remains displayed until the
next measurement or the rotation of the switch.
If a voltage higher than the authorised limit value (see § 3.2) appears during all the measurements, the
device will indicate this voltage on the small display using the fl ashing warning symbol and stop the
measurement in progress.
Note: A certain number of special functions may be used (see § 4).

5.2. INSULATION MEASUREMENT

(See § 3.2)
With this function, the instrument can measure insulation between 10 k  and 10 T  , according to the
chosen test voltage (among 500 - 1000 - 2500 - 5000 V) or programmed voltage (between 40 V and 5100 V).
 Position the switch on "500 V-2 T  ", or "1000 V-4 T  ", or "2500 V-10 T  ", or "5000 V-10 T  " or
"Var 50-5000 V"
 Connect the device to the part to be tested.
If the voltage present is higher than the authorised limit value (see § 3.2) the measurement
will be inhibited.
 Start the measurement and note the results.
It is possible to scroll through all the results on the main display using the R-DAR-PI-DD key (see § 4.3)
or on the small display using the V-TIME key (see § 4.2).
R(t) is used to access the intermediate values measured and stored at a rate set in SET-UP, in
"Programmed-time test" mode. These samples are available until the start of another measurement or
until the next rotation of the switch (see § 4.3)
To measure high insulation values (> 1 G  ), you are advised to use the "G" guard terminal to remove
the infl uence of surface leakage currents. The guard terminal should be connected to a surface where
surface leakage currents may fl ow via dust and humidity: for example, the insulation surface of a cable
or transformer, between two measurement points.
As soon as insulation measurements stop, the tested circuit is automatically discharged via an
internal resistor of the device.

5. USE

62

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

C.a 6547

Table des Matières