4.4 Overall movements
All movement of the lamp must be done through the handles of the head.
The arm of the lamp can be moved according with following schemes:
!
Do not move the lamp or the head with the switch lever.
Do not move the lamp or the head outside the limits described above.
4.5 Acoustic Warning Signals
OTP* = Beep 30 seconds
MAX = 1 beep - 2 beep with proximity
MIN = 1 beep
1 BEEP = On the commands
1 BEEP = At start-up
OpL** = Beep 30 seconds
* OTP: Over-temperature protection
** OpL: LED charger unplugged
5. MAinTenAnce And PeRiOdic cHecKS
The maintenance of the lamp is made up by following operations:
Check the absence of space between joint components of the arms (yearly)
Check readability of data on the information plate (yearly)
Electrical safety check according to EN 62353 (every two years)
1. Insulation resistance
2. leakage current
Light checks (every five years or 10,000 hours of function)
1. Maximum lighting intensity >35000 lux
2. CRI decay: <20% for data on the information plate.
3. Expected value of Blue light on spectrum emitted, measured in w/m2: <100.
Fig 3
21