3. Calibrations

3.1 Calibration of the unit

The calibration of the unit consists in entering 2 reference points on the probe's
measuring range and specify the distance between them. This is to be repeated for
both channels if required.
The unit is factory calibrated with a probe master. If the unit is not calibrated, the
calibrated function is automatically called when powering the unit.
+
Prob = 1
rEF0 CH1
....
rEF0 CH1
010.0000
Second part of the diagram, see next page
For a new calibration, press Preset key when switching ON
the unit. (simultaneously).
Prob = 2
Calibration of channel 1 or
Calibration of channel 2 or
Calibration of 2 channels (1 and 2)
Place the probe on reference 0
Acquisition of the reference 0
Insert the reference value (slip gauge value) using
The key
and
reference 1, then press Preset key.
47
Prob =1,2
then place the probe on
E