The B-Max Plus curing light is plated with metal. After each use, moisten a gauze or soft
cloth with an anti-microbial surface disinfectant and wipe the surface and lens.
Periodically check the lens for cured dental resins. lf necessary, use a non-diamond
instrument to carefully remove any adhered resin.
The barrier sleeve prevents cross contamination and helps keep dental composite material
from adhering to the surface of the lens and wand body.
CAUTION
-Do not autoclave batteries, charger, power adapter or B-Max Wand.
-Do not insert fingers, instruments or other objects into the battery compartment
of the Wand.
-Do not attempt to clean the gold contacts or any part of the battery compartment.
-Do not immerse in any kind of ultrasonic bath or any liquids.
-Do not wipe down B-Max Plus curing light with caustic or abrasive cleaners.
18
MAINTENANCE AND CLEANING