Test frequency : 10 kHz
Range
1000 mH
200 mH
20 mH
2000 µH
Notes
IX 3131B
Max.
display
999.9 mH
199.99 mH
19.999 mH
1999.9 µH
1 -
Q Value is the reciprocal of DF.
2 -
This specification is based on the measurement performed
at the test leads.
4 -
Lx = counts of displayed L value. e.g. L = 88.88H, then
Lx = 8888.
5 -
The accuracy is specified for series mode in this range.
Accuracy (DF < 0.5)
Inductance
Dissipation Factor
2.0 % +
(Lx/10000) % + 8
1.5%+100/Lx+10
*N4
0.6 % +
1.5%+100/Lx+10
(Lx/10000) % + 8
0.6 % +
2.0%+100/Lx+15
(Lx/10000)%+10
1.0 % +
5.0%+100/Lx+20
(Lx /10000)% +10