Application Parameters; Service Life Of Accessories - StarLab Microcentrifuge 24 Instructions D'origine

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Technical data
Microcentrifuge 24
9.5

Application parameters

Tab. 9-1: Acceleration time and braking time according to DIN 58 970
Rotor
Standard 24 AT
PCR 4x8
Cycle time
Rotational speed
Relative centrifugal force
Maximum load
Maximum kinetic energy
Permitted density of the material for centrifuging
(at maximum g-force (rcf) or rotational speed (rpm)
and maximum load)
Inspection obligation in Germany
9.6

Service life of accessories

CAUTION! Danger due to material fatigue.
When the service life is exceeded, it cannot be guaranteed that the material of the rotors and accessories will withstand the stresses
during centrifugation.
 Do not use any accessories which have exceeded their maximum service life.
All rotors and rotor lids can be used during the entire service life of the centrifuge if the following
conditions are met:
• proper use
• recommended maintenance
• undamaged condition
Accessories
Aerosol-tight rotor lid
Seals in the aerosol-tight rotor lid
Adapter
The date of manufacture is stamped on the rotors in the format 2015-03 (= March 2015).
30
· Microcentrifuge 24 · Manual
Acceleration time
15 s
15 s
Maximum service life after initial setup
50 autoclaving cycles
Deceleration time
15 s
15 s
10 s – 9:59 h, unlimited (∞)
• 10 s – 2 min: can be set in increments of 10 s
• 2 min – 10 min: can be set in increments of 30 s
• 10 min – 9:59 h: can be set in increments of 1 min
100 rpm – 15060 rpm
• 100 rpm – 5000 rpm: can be set in increments of 10 rpm
• 5000 rpm – 15060 rpm: can be set in increments of 100 rpm
10 × g – 21300 × g
• 10 × g – 3000 × g: can be set in increments of 10 × g
• 5000 × g – 21300 × g: can be set in increments of 100 × g
Fixed-angle rotor: 24 × 2 ml
4.12 kJ
1.2 g/ml
no
3 years
1 year