4.2
Preparing installation
The weight of the centrifuge is 12.98 kg.
Unpacking the centrifuge
1. Open the packaging board.
2. Remove the accessories.
3. Lift the centrifuge out of the carton.
4. Place the centrifuge on a suitable lab bench.
5. Remove the plastic sleeve.
The device is delivered with and without a rotor.
The transport securing device can be removed directly from devices without a rotor.
6. Turn the rotor nut counterclockwise using the supplied rotor key.
7. Lift the rotor out vertically.
8. Remove the transport securing device.
4.3
Installing the instrument
Prerequisites
The device is on a suitable lab bench.
WARNING! Danger due to incorrect voltage supply.
Only connect the device to voltage sources which correspond with the electrical requirements on the name plate.
Only use earth/grounded sockets with a protective earth (PE) conductor.
Only use the mains/power cord supplied.
NOTICE! Damage to electronic components due to condensation.
Condensate may form in the device when it has been transported from a cool environment to a warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains/power line.
1. Let the device warm up to ambient temperature.
2. Connect the centrifuge to the mains and switch it on using the mains/power switch.
• The RPM/RCF key lights up.
• The display is active.
Installation
Microcentrifuge 24
15
Microcentrifuge 24 · Manual ·