12 General algorithm
1
No electrodes
fault detected
PRESS ANALYSE KEY (Rem 1)
Xð 2
q 5 seconds
Analyse key
pressed
10 s time out
Note :
q 20 seconds
= Text 1 repeated every 20 seconds
Xð1
Rem 1 : The step "PRESS ANALYSE" (text 2) is cancelled if the analyse is automatic or if the
device functions in the automatic mode.
Rem 2 : The text 14 replaces the text 5 if the device functions in the automatic mode.
SAED Power on
ELECTRODES FAULT
Xð 1
ANALYSE IN PROGRESS
Non shockable rhythm
NO SHOCK ADVISED
Xð 6
Non shockable rhythm
RYTHM HAS CHANGED
3 s time out
3 s time out
Xð 8
2 minutes passed
&
CPR 30:2
q 20 seconds
ANALYSE
X ð 3
Xð 7
Shock key pressed
CPR 30:2
q 1 minute
1
32
Electrodes fault
detected
Shockable rhythm
SHOCK ADVISED
Xð 4
Defi charged
GIVE 1 SHOCK (Rem 2)
Xð 5 or 14
q 10 seconds
Internal
discharge
Time out